Using our unique patented sensor platform, the sensors are more robust than traditional technologies and require no reagents.
An integral stirrer maintains flow around the sensors, eliminating flow to waste.
Operational costs are less than chemical alternatives.
Compact and lightweight in construction, it is easy to install in a boundary box when a water meter is removed.
Easy to Install
The sensor head is mounted in a plastic manifold allowing the water to pass over the sensors. The water company will remove an existing water meter from the boundary box and the SA is temporarily fitted in its place via a 1.5" BSP fitting.
A battery pack is connected to the enclosure.
Sensor Performance
The Chlorine, Mono-chlorines and Dissolved Oxygen Sensors use microelectrode technology for maximum life.
The sensors are membrane free and insensitive to pressure and flow. They have a fast response, typically <20seconds, 95% of step change.
Dimensions and Weight
• 30cm total height
• 3.5kg approx
Power Requirements
• Battery Pack 12v (8C Cells)
Intellisonde SA Specification
| Parameter | Performance | |
|---|---|---|
| Temperature | -5 to +50˚C ± 0.2˚C | |
| Colour | 0 to 50 Hazen ± 1 Hazen | |
| PH (Glass) | 2 to 12pH ± 0.2pH in typical installations | |
| Dissolved Oxyen | 0 to 20 mg/l ± 10% | |
| ISE | 0 to 1 volt ± 1 mV | |
| Free Chlorine & Mono-Chloramines (Total Chlorine) | 0 to 5 mg/l ±10% or ± 0.05 mg/l | |
| Free Chlorine | 0 to 5 mg/l ±10% or 0.05 mg/l | |
| Pressure(standard) | 0 to 10 Bar | |
| Turbidity | 0 to 50 NTU ± 0.5 NTU | |
| Conductivity | 0 to 10 mS/cm ± 5% | |
| ORP (REDOX) | +1.0 volt to -1.0 volt ± 1mV | |